Mass Spectrometry Sampling Under Ambient Conditions with Desorption Electrospray Ionization

See allHide authors and affiliations

Science  15 Oct 2004:
Vol. 306, Issue 5695, pp. 471-473
DOI: 10.1126/science.1104404

Abstract

A new method of desorption ionization is described and applied to the ionization of various compounds, including peptides and proteins present on metal, polymer, and mineral surfaces. Desorption electrospray ionization (DESI) is carried out by directing electrosprayed charged droplets and ions of solvent onto the surface to be analyzed. The impact of the charged particles on the surface produces gaseous ions of material originally present on the surface. The resulting mass spectra are similar to normal ESI mass spectra in that they show mainly singly or multiply charged molecular ions of the analytes. The DESI phenomenon was observed both in the case of conductive and insulator surfaces and for compounds ranging from nonpolar small molecules such as lycopene, the alkaloid coniceine, and small drugs, through polar compounds such as peptides and proteins. Changes in the solution that is sprayed can be used to selectively ionize particular compounds, including those in biological matrices. In vivo analysis is demonstrated.

An awkward feature of mass spectrometry (MS) is that the sample must be introduced into vacuum or into an inaccessible region closely coupled to the vacuum system. Here we describe a simple approach that allows ambient sampling for MS analysis. Electrosprayed (ES) aqueous droplets are directed at a surface of interest in air. The sample can be moved continuously or reoriented in space while MS analysis proceeds. The microdroplets act as projectiles and desorb ions from the surface as a result of electrostatic and pneumatic forces (1). The desorbed gas-phase ions are transferred to the distant mass spectrometer via an atmospheric pressure ion-transfer line.

This new method, termed desorption electrospray ionization (DESI), is related to other spray ionization methods, including electrospray ionization (ESI) (2, 3), and to the desorption ionization methods, such as secondary ion mass spectrometry (SIMS) and laser desorption (4, 5). No matrix is needed to perform the experiment—an advantage shared with laser desorption from porous silicon surfaces (6)—and the production of multiply charged biological ions is advantageous in extending the mass range, as is the case with ESI.

In its simplest form, the desorption electrospray experiment (Fig. 1) uses an aqueous spray directed at an insulating sample or an analyte deposited on an insulating surface such as polytetrafluoroethylene (PTFE). The desorbed ions are sampled with a commercial ion trap mass spectrometer equipped with an atmospheric interface connected via an extended and preferably flexible iontransfer line made either of metal or an insulator.

Fig. 1.

Schematic of typical DESI experiment. The sample solution was deposited from solution and dried onto a PTFE surface, and methanol-water (1:1 containing 1% acetic acid or 0.1% aqueous acetic acid solution) was sprayed at a flow rate of 3 to 15 μl/min under the influence of a high (4 kV) voltage. The nominal linear velocity of the nebulizing gas was set to 350 m/s.

Examination of insulator surfaces is highly unusual in MS; however, the fast nebulizing gas jet in the DESI experiment transports the charged microdroplets and allows them to impact the surface and to carry away analyte molecules. The underlying processes might be related to those occurring in versions of SIMS that use clusters as projectiles (711) or impact desolvation of droplets on surfaces (12) in vacuum environments.

A broad range of analytes have been examined, from simple amino acids through drug molecules, alkaloids, terpenoids, and steroids, to peptides and proteins. The methodology seems to be particularly promising for forensic and public-safety applications, including analysis of dried blood, detection of explosives, and monitoring of chemical warfare agents (13), as illustrated by two experiments. In one experiment, the explosive RDX was desorbed from an insulating tanned leather surface, to give the mass spectrum shown in Fig. 2A. In the other experiment, nitrile gloves that were exposed for less than 1 s to dimethyl methylphosphonate vapors (DMMP, a chemical warfare agent simulant), and then washed and dried, gave a mass spectrum (Fig. 2B) that unequivocally indicates the presence of trace levels of DMMP.

Fig. 2.

(A) Negativeion DESI spectrum of RDX (1 ng/mm2) on leather (porcine) surface using acetonitrile/methanol/trifluoroacetic acid (TFA) 1:1:0.1% as spray solvent. The presence of the explosive is confirmed by tandem MS (inset). (B) Positive-ion DESI spectrum of DMMP, a chemical warfare agent simulant, from a nitrile glove surface using the same spray.

New applications of MS might emerge from such simple sampling procedures. In particular, process analysis and other high-throughput experiments are much simplified over standard MS methods. Initial experiments with pharmaceuticals show analysis rates of 20 samples per second (1). Optimum experimental conditions are summarized in table S1 (1). The ultimate sensitivity of the DESI method has not been determined, but lysozyme present in amounts ranging from 10 to 50 pg could be detected.

A feature of DESI relative to traditional desorption ionization methods, and indeed other MS methods, is the ease with which chemical reagents can be supplied to the site of analysis. This allows the generation of specific reaction products that can be used to confirm analyte identification. Biochemical reactions, including the formation of noncovalent complexes between enzymes and substrates, also serve this purpose. For example, when the lysozyme substrate hexa-(N-acetyl) chitohexaose was included in the solution sprayed onto a lysozyme sample, the enzyme-substrate complex was seen at mass-to-charge (m/z) ratios of 1944 and 2220 (fig. S3) (1, 1416). Another example of a specific chemical reaction used to confirm MS identification is the formation of metal complexes between an analyte on the surface and a metal ion introduced into the spray solution. Uses for this capability include experiments (1) in which the chirality of amino acids is measured via diastereomeric complex ion formation and fragmentation (17, 18).

Both MALDI (matrix-assisted laser desorption/ionization) (1921) and SIMS (5, 22, 23) can be used to image biological materials in experiments usually done in vacuum. The exceptions are atmospheric pressure (AP)–MALDI (21, 24) and AP-laser ablation (25), but in both of these methods the sample is strictly positioned relative to the rest of the ion source and is inaccessible and not manipulated during the experiment. Working under ambient conditions, DESI can be used for the spatial analysis of native surfaces, such as plant or animal tissues. The potential for this type of application is illustrated by the DESI spectrum of a seed section of poison hemlock (Conium maculatum) (Fig. 3A). The peak at m/z 126 is due to coniceine, which is known to be present in this particular plant species. The possibility of in situ imaging was demonstrated by scanning the spray spot across a cross section of the plant stem (Fig. 3B). Similarly, the DESI spectrum collected from tomato (Lycopersicon esculentum) skin also indicates the localization of characteristic compounds, including lycopene at m/z 536 (Fig. 3C).

Fig. 3.

(A) Positive-ion DESI spectrum of Conium maculatum seed section with the sample held under ambient conditions. The signal at m/z 126 corresponds to protonatedγ-coniceine (molecular weight 125), an alkaloid present in the plant. The inset shows the MS/MS spectrum of m/z 126. (B) The intensity distribution of m/z 126 across a stem cross section. (C) DESI mass spectrum of the tomato shown in photograph S2. (1) The peak at m/z 536 is due to lycopene and/or other carotenoids. Methanol-water was sprayed onto the tomato surface and desorbed ions were transferred to the ion trap mass spectrometer.

Quantitative results can be obtained by using appropriate internal standards in experiments where the sample is deposited on a target surface; however, quantification by any method is intrinsically difficult in the analysis of natural surfaces. Sprayed compounds used as internal standards yielded semiquantitative results (relative standard deviation values of ∼30%) for spiked plant tissue surfaces.

We have also used DESI for in vivo sampling of living tissue surfaces. An aqueous-alcohol DESI spray was directed onto the finger of a person who had taken 10 mg of the over-the-counter antihistamine Loratadine. About 40 min after taking the tablet, the molecule became detectable directly on the skin or in saliva and its surface concentration remained above the detection limit for another 50 min. A typical spectrum taken directly from skin (photograph S3) is shown in Fig. 4. This example is representative of the novel applications of MS once its vacuum constraints are lifted.

Fig. 4.

DESI spectrum recorded by spraying methanol-water onto the finger of a person 50 min after taking 10 mg of the over-the-counter antihistamine Loratadine (m/z 383/385). Other peaks, e.g., at m/z 245, are either due to the lab air background or are associated with the detergent that was used for hand washing before the experiment. Note that mass assignments are from the instrument data system and are accurate to just 0.5 m/z units.

Supporting Online Material

www.sciencemag.org/cgi/content/full/306/5695/471/DC1

Materials and Methods

Table S1 Figs. S1 to S8

Photographs S1 to S3

References and Notes